Criticare nGenuity™ 8100 Monitor conscious sedation or post-procedure recovery monitoring
Criticare nGenuity™ 8100E Monitor Conscious Sedation or Post-Procedure Recovery Monitoring

The Criticare nGenuity 8100E Series patient monitor is ideal for monitoring pulse oximetry (SpO2), non-invasive blood pressure, heart rate, temperature, respiration and ECG during conscious sedation or post-procedure recovery. It comes with a full compatibility with popular Electronic Medical Records (EMR) systems makes it easy for any facility or office to transfer patient data. Lightweight design and has a carry handle, the nGenuity 8100E monitor also can be wall mounted or attached to a rolling stand for increase any medical setting. For portability, a sealed battery inside the unit can be recharged with a standard AC plug.
Features:
CAPNOGRAPHY (CO2)

CAPNOGRAPHY (CO2)
Units: |
mmHg; kPa; Torr; Percent |
Method: |
Non-dispersive infrared; autocalibrating |
N2O Compensation: |
Manual (On/Off, user selectable |
Range: |
0 to 99 mmHg; 0 to 12.5%; |
Flow, Sample Rate: |
Sidestream; 200 ml/min |
Accuracy: |
±2 mmHg, ±0.3 volume %, ±0.3kPa, or ±2 Torr at 200 ml/min and respiratory rate equal or less than 120 bpm |
Rise Time: |
170 ms (10 to 90%) at 200 ml/min |
ECG 0r CG or EKG


ECG
Number of Leads: |
3 – or 5 Lead |
Lead View: |
User-selectable; I, II, III, aVR, aVL, aVF, V (5-Lead); I, II, or III (3-Lead) |
Gain Selection: |
x0.5, x1, x2, and x4 |
Frequency Response: |
Diagnostic: 0.05 to 100 Hz (-3dB); |
Monitor: |
0.5 to 40 Hz (-3dB); |
ST: |
0.05 to 40 Hz (-3dB |
Electrosurgery Protection: |
Yes |
Defibrillator Protection: |
Yes |
Pacer Detection/Rejection: |
Yes |
ST Accuracy: |
±0.2 mm or ±5% of reading, whichever is greater, over heart range of 30 to 120 bpm |
ST Range: |
-6.0 to 6.0 mm |
|
|
Non-Invasive Blood Pressure (NIBP)

NIBP
Technique: |
Oscillometric measure upon inflation |
Inflation Range: |
Adult/Pediatric: 0 to 300 mmHg: Neonate: 0 to 150 mmHg |
Measurement Cycle: |
< 40 sec. typical; standard adult cuff |
Auto. Measurement: |
2, 3, 5, 10, 15, 30 min; 1, 2, 4 hour intervals |
STAT Mode: |
5 minutes of consecutive readings |
Maximum Cuff Pressure: |
Adult/Pediatric: 300 mmHg; Neonate: 150 mmHg |
Resolution: |
1 mmHG |
Transducer Accuracy: |
±2 mmHg or ±2% of reading, whichever is greater, over full range |
Clinical Accuracy: |
Complies with AAMI |
PULSE OXIMETRY

Range: |
1% to 99% |
Resolution: |
1% |
Accuracy: |
70% to 99% range: ±2%; 50% to 69% range: ±3%; < 50%: unspecified |
Method: |
Dual wavelength LED |
HEART RATE (PULSE)
Source: |
User selectable: Smart, ECG, Pleth, NIBP |
Range: |
20 to 300 bpm (ECG, Pleth); 30 to 240 bpm (NIBP) |
Accuracy: |
±1 bpm or ±1% (ECG) whichever is greater |
TEMPERATURE
Channels: |
1 |
Probe: |
YSI-400 and YSI-700 Series |
Range, Accuracy: |
68° to 113°F (20° to 45°C): ±0.1°C |
Display Resolution: |
0.1°C |
RESPIRATION
Rate |
6 to 150 bpm (ECG); 2 to 120 (CO2) |
Resolution: |
1 breath/min |
Accuracy: |
±1 breath/min |
Source: |
User selectable: Smart, CO2, ECG-TTI |